Device product
Brilliance 64 Computed Tomography X-ray system
Z-0544-2016
Product summary
- Event
- Event 72684
- Status
- Terminated
- Classification
- Class II
- Quantity
- 30 units
- Official record key
device-enforcement:Z-0544-2016
Official wording
Reason: Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.
Code information: Model No. 728231; S/N: 4001, 9074, 9099, 9538, 9549, 9573, 9638, 9646, 9656, 9667, 9703, 9738, 9755, 9782, 9811, 9822, 9909, 9923, 9927, 9933, 10021, 10042, 10077, 10150, 10180, 10379, 29056, 29107, 29115, 90124, 90154, 95024, 95130, 95222, 95236, 95258, 95261, 95364, 95399, 95410, 95475, 95551, 95614, 95668, 95688 & 95691.
Distribution pattern: Worldwide distribution. US nationwide, Australia, Austria, Bangladesh, Belgium, Brazil, Cambodia, Canada, China, Egypt, France, Germany, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Lebanon, Libya, Lithuania, Malaysia, Martinique, Myanmar, Mexico, Mozambique, Netherlands, New Zealand, Norway, Palestine, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Tanzania, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uzbekistan & Vietnam.
Derived failure modes
-
Device software or design
software (versions preceding 4.1.2) without buying the helical perfusion license key could