Skip to content
Recall Observatory FDA recall evidence

Device product

EDI BZO Uncut Sheet - Cup*, Item No. 160170

Z-0501-2016

November 16, 2015

Class III

Product summary

Firm
Ameditech Inc
Event
Event 72651
Status
Terminated
Classification
Class III
Quantity
45,584 kits
Official record key
device-enforcement:Z-0501-2016

Official wording

Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

Code information: 141260 141294 141909 141925 142074 142243 142428 142614 142712 142911 144555 151925 152003 152223 152498 152723 152897 153311 153559 153818 154184 152103

Distribution pattern: Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.