Cable remote control JUPITER / TS7500, Material no.1601926, Serial number: to (incl.) 101465919. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Z-0919-2015

December 03, 2014

Class II

Product summary

Firm
Trumpf Medical Systems, Inc.
Event
Event 69927
Status
Terminated
Classification
Class II
Quantity
3303 total all devices
Official record key
device-enforcement:Z-0919-2015

Official wording

Reason: Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes

Code information: The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.

Distribution pattern: Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes