Device product
SYS IVK, Bio mCT-S(40) 3R->4R Upgrade, MATERIAL NUMBER 10250743 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Z-2035-2016
Product summary
- Event
- Event 74192
- Status
- Terminated
- Classification
- Class II
- Quantity
- 4 units
- Official record key
device-enforcement:Z-2035-2016
Official wording
Reason: Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
Code information: Serial Numbers: 1002,1001,11003,11004
Distribution pattern: Worldwide Distribution - US Nationwide
Derived failure modes
-
Unknown
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.