Device product
CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.
Z-0001-2017
Product summary
- Event
- Event 75118
- Status
- Terminated
- Classification
- Class III
- Quantity
- 140,000 units
- Official record key
device-enforcement:Z-0001-2017
Official wording
Reason: Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.
Code information: 6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various
Distribution pattern: Worldwide Distribution - USA (nationwide) Distribution.
Derived failure modes
-
Unknown
Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.