CDI¿ H/S Cuvette The intended use for the CDI¿ H IS Cuvette is as follows: The CDI¿ H/S Cuvettes are intended for use with a CDI monitor during cardiopulmonary bypass procedures when continuous monitoring of blood hematocrit and oxygen saturation is desired.

Z-0001-2017

December 11, 2015

Class III

Product summary

Firm
Terumo Cardiovascular Systems Corporation
Event
Event 75118
Status
Terminated
Classification
Class III
Quantity
140,000 units
Official record key
device-enforcement:Z-0001-2017

Official wording

Reason: Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.

Code information: 6912 , 6913 6914, 6932, 6933 6934, 6922, 6923 6924, CV-6912, CV-6913, CV-6914 Cardiovascular Procedure Kits- Various

Distribution pattern: Worldwide Distribution - USA (nationwide) Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.