Extended Brilliance Workspace NM Nuclear medicine image display and processing application suite. AutoSPECT Pro Software Application. Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.

Z-1504-2016

March 04, 2016

Class II

Product summary

Firm
Philips Medical Systems (Cleveland) Inc
Event
Event 73776
Status
Terminated
Classification
Class II
Quantity
126 Units
Official record key
device-enforcement:Z-1504-2016

Official wording

Reason: The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

Code information: M/N 882489; S/N: 66794, 66798, 66836, 66009,66011, 67154, 66828, 67058, 67050, 67031, 67284, 66332, 66335, 66397, 67026, 67040, 66084, 66089, 66092, 67007, 67011, 67226, 66956, 66454, 66447, 66413, 66129, 66128, 66300, 66023, 11728, 66000, 66938, 67270, 67252, 66571, 66443, 66324, 66869, 68477, 66682, 66815, 66284, 67263, 66265, 66277, 66880, 66892, 66258, 66079, 66181, 66451, 66627, 66187, 66806, 67016, 67008, 66706, 66745, 66817, 66822, 66910, 66908, 66811, 67087, 67094, 66241, 66825, 66632, 66624, 66854, 66689, 66690, 66906, 66792, 67173, 67069, 67166, 68290, 67182, 66785, 67292, 67286, 66768, 66769, 66550, 66113, 67220, 67193, 67186, 66657, 66668, 66564, 66065, 67065, 67208, 67204, 66818, 66819, 67122, 66818, 66819, 66456, 67212, 66019, 67057, 67172, 67157, 67167, 67162, 66737, 68361, 66183, 66368, 66387, 66376, 66415, 66136, 66137, 66960, 66958, 67142 & 67165.

Distribution pattern: Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.