SDMS (SenTec Digital Monitoring System)

Z-0088-2017

September 15, 2016

Class II

Product summary

Firm
SenTec AG
Event
Event 75257
Status
Terminated
Classification
Class II
Quantity
75 (54 US; 21 =OUS)
Official record key
device-enforcement:Z-0088-2017

Official wording

Reason: Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).

Code information: SN 306497 to SN 306571.

Distribution pattern: Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).