TBS iNsight Version v.3.0.1 Product Usage: TBS iNsight is a medical device software that is installed on bone densitometers for analysis of bone microarchitecture and osteoporosis management.

Z-0369-2017

October 03, 2016

Class II

Product summary

Firm
Medimaps Group
Event
Event 75510
Status
Terminated
Classification
Class II
Quantity
15 units in US
Official record key
device-enforcement:Z-0369-2017

Official wording

Reason: The FRAX adjusted for TBS values are not correct when: The FRAX feature is activated in TBS iNsight; and TBS has been computed from a spine scan where some vertebrae were excluded.

Code information: Version v.3.0.1; Serial # 14295, 300409, 70580, 200042, 500581, 70103, 70174, 80936, 100202, 201120, 84584, 83769, 100112, 210020, 210440

Distribution pattern: Worldwide Distribution - US Nationwide - in the states of: IL, NC, CA, DE, NY, MD, WI, MN. And the following countries: France,Thailand, Croatia, Canada, Italy, Spain, Russian Federation, Ukraine, Lithuania, South Africa, Brazil, Lebanon, Saudi Arabia, Korea, China Portugalm Iraq, Syrian Arab Republic, Czech Republic, Australia, Belgium, Austria, Romania, New Zealand, Switzerland, Hong Kong, Sweden, Mexico, Islamic Republic of Iran, Serbia, Netherlands, Denmark, and UK.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The FRAX adjusted for TBS values are not correct when: The FRAX feature is activated in TBS iNsight; and TBS has been computed from a spine scan where some vertebrae were excluded.