Device product
Fresenius Liberty Cycler-Peritoneal Dialysis Cycler Model: 180111 and RTLR180111
Z-0693-2016
Product summary
- Event
- Event 72926
- Status
- Terminated
- Classification
- Class II
- Quantity
- 14,094 units
- Official record key
device-enforcement:Z-0693-2016
Official wording
Reason: The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).
Code information: All lot numbers
Distribution pattern: Nationwide Distribution
Derived failure modes
-
Unknown
The door latch may not fully close and the door may open unexpectedly. The door may make physical contact with anything near the door of the machine, including the patient. May cause injury, delay in treatment, or Increased Intraperitoneal Volume (IIPV) (overfill).