Device product
Algostim LLC, Algovita External Pulse Generator, Trial Stimulator, Model 3400
Z-1194-2016
Product summary
- Firm
- QiG Group LLC
- Event
- Event 73150
- Status
- Terminated
- Classification
- Class II
- Quantity
- 31
- Official record key
device-enforcement:Z-1194-2016
Official wording
Reason: The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.
Code information: Serial Numbers: 01DC94, 01DC98, 01DC9B, 01DCA6, 01DCB2, 01DCB3, 01DCB4, 01DCB5, 01DCB6, 01DCB
Distribution pattern: Distributed in Germany and Luxembourg.
Derived failure modes
-
Unknown
The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.