Device product

STAT-Check Resuscitation Bag Manual Emergency Resuscitation Bag with Carbon dioxide gas analyzer Intended for use in various clinical situations to provide respiratory support in the presence of reversible apnea commonly associated with respiratory arrest. The device should only be used by personnel trained in CPR procedures. The ACE/STAT-Check II is intended for use as an adjunct assessment tool for verification of successful ventilation / intubation in patients > 2.2 kilograms (5lbs.).

Z-1880-2016

March 15, 2016

Class II

Product summary

Firm
Ventlab LLC
Event
Event 73766
Status
Terminated
Classification
Class II
Quantity
2430
Official record key
device-enforcement:Z-1880-2016

Official wording

Reason: Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.

Code information: Model SC9001B Adult STAT-Check Resuscitation Bag w/ Inflatable Bag Reservoir & Mask Lot numbers 303320 and 303612 Model SC9001C Adult STAT-Check Resuscitation Bag w/ Corrugate Tube Reservoir & Mask Lot number 303400 Model SC9001C-C Adult STAT-Check Resuscitation Bag w/ Corrugate Tube Reservoir, Mask & Straight Connector Lot numbers 303523 and 303791 Model SC8001B Pediatric STAT-Check Resuscitation Bag w/ Inflatable Bag Reservoir & Mask Lot number 303833

Distribution pattern: Nationwide distribution to AZ, MS, NY, MO, FL, WI, TX, FL, and CA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during use and will not be an indicator of CO2 level or improper performance of cardiac message and ventilation.