Device product
Merge Eye Station Import Utility (ESIU) when used with Merge Eye Station and Merge Eye Care PACS. The firm name on the labeling is Merge Healthcare, Hartland, WI.
Z-1246-2017
Product summary
- Event
- Event 75914
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1,627 sites potentially have the affected versions
- Official record key
device-enforcement:Z-1246-2017
Official wording
Reason: Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN)
Code information: Merge ESIU: Versions 1.0, 1.1.1, 1.1.2, and 1.1.3; Merge Eye Station: All versions; Merge Eye Care PACS: Versions 4.1, 4.2, 4.3, 5.0, and 5.1
Distribution pattern: Distribution was nationwide to medical facilities. There was also government and foreign distribution, including Canada.
Derived failure modes
-
Unknown
Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN)