Device product
Salto Talaris Tibial Guide, Part # MJU333 ,Tray/Kit Code YKAL11, Non sterile. Orthopedic Manual Surgical Instrument
Z-2465-2015
Product summary
- Firm
- Tornier, Inc
- Event
- Event 71913
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5
- Official record key
device-enforcement:Z-2465-2015
Official wording
Reason: Tornier has initiated a voluntary recall for the Salto Talaris Tibial Guide (part number MJU333) due to a complaint that occurred during physician training, that the guide cannot be assembled to the Salto Talaris Talar Pin Setting guide. This issue will cause the impossibility the time of putting the Talar pin in the guide.
Code information: Lot #14F005
Distribution pattern: Distributed to CA, ME, NC, PA, and TX.
Derived failure modes
-
Unknown
Tornier has initiated a voluntary recall for the Salto Talaris Tibial Guide (part number MJU333) due to a complaint that occurred during physician training, that the guide cannot be assembled to the Salto Talaris Talar Pin Setting guide. This issue will cause the impossibility the time of putting the Talar pin in the guide.