Device product
ABG-HM-1 Hummi Micro Draw Blood Transfer Device Product Usage: It is used for blood transfer and collection from Peripheral Arterial Line Catheters.
Z-1636-2016
Product summary
- Firm
- Hummingbird Med
- Event
- Event 72350
- Status
- Terminated
- Classification
- Class I
- Quantity
- 11500 units
- Official record key
device-enforcement:Z-1636-2016
Official wording
Reason: Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow cannula hub may not be sufficiently secure enough and could separate while in use.
Code information: Lot 15180, 15286, 15287, 15300, 15305
Distribution pattern: Distributed in the states of CA, KY, MD, GA, and IL.
Derived failure modes
-
Unknown
Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow cannula hub may not be sufficiently secure enough and could separate while in use.