Device product
Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION INSTRUMENTS REF 6543-1-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA A Subsidiary of Stryker Corp. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France The Offset Bushings are parts of the Triathlon TS system. They are used during trialing to assess the appropriate magnitude and orientation of offset between the Implant and the intramedullary Stem. The feedback from the Offset Bushing allows for the selection of the corresponding Offset Adaptor, which allows for optimal bone coverage, compensating for differences between the Implant and bone canal axes. Made in USA
Z-0469-2013
Product summary
- Event
- Event 63442
- Status
- Terminated
- Classification
- Class II
- Quantity
- 7 units
- Official record key
device-enforcement:Z-0469-2013
Official wording
Reason: Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.
Code information: Class I, 510(k) exempt Catalog Number: 6543-1-600 Lot Number; NE1WR1
Distribution pattern: Worldwide Distribution.
Derived failure modes
-
Unknown
Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.