Skip to content
Recall Observatory FDA recall evidence

Device product

Catalog Number: 900741A ORTHO-IMPLANT SURGICAL PACK Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Z-1485-2016

January 07, 2016

Class II

Product summary

Firm
Customed, Inc
Event
Event 72992
Status
Terminated
Classification
Class II
Quantity
153
Official record key
device-enforcement:Z-1485-2016

Official wording

Reason: Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.

Code information: LOTS: SL15174 (150619406) 30-JUN-16 SL15197 (150719810) 31-JUL-16 SL15201-02 (150719810) 31-JUL-16 SL15231 (150820049) 31-AUG-16 SL15232 (150820049) 31-AUG-16 SL15243 (150820049) 31-AUG-16 SL15265 (150820558) 31-AUG-16 SL15268 (150920711) 30-SEP-16

Distribution pattern: Distributed only in Puerto Rico.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    inadequate validation/qualification of sterilization processes
  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    inadequate validation/qualification of sterilization