Device product
Immutest 5 Pnl Drug Cup w/ Adult -waived, Item No. IMCA-5AB-W
Z-0509-2016
Product summary
- Firm
- Ameditech Inc
- Event
- Event 72651
- Status
- Terminated
- Classification
- Class III
- Quantity
- 1,909 kits
- Official record key
device-enforcement:Z-0509-2016
Official wording
Reason: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Code information: 141758 141921 142657 143178 143581 144563 151337 152240 152782 153154 153401
Distribution pattern: Nationwide Distribution.
Derived failure modes
-
Unknown
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.