Device product
VITROS Chemistry Products Performance Verifier is used as an assayed control to monitor the performance on VITROS 250/350/5, 1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Used as a control for the VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Z-0807-2018
Product summary
- Event
- Event 78724
- Status
- Terminated
- Classification
- Class III
- Quantity
- 10,021 units
- Official record key
device-enforcement:Z-0807-2018
Official wording
Reason: VITROS Performance Verifier II (PVII) Lot K4852 does not meet long term stability requirements for the AST Analyte (only) when the product is stored refrigerated for up to 6 months.
Code information: Lot K4852, Expiration 22 Feb 2018
Distribution pattern: Worldwide
Derived failure modes
-
Unknown
VITROS Performance Verifier II (PVII) Lot K4852 does not meet long term stability requirements for the AST Analyte (only) when the product is stored refrigerated for up to 6 months.