Skip to content
Recall Observatory FDA recall evidence

Device product

Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Z-2097-2017

April 10, 2017

Class II

Product summary

Firm
Ion Beam Applications S.A.
Event
Event 76977
Status
Terminated
Classification
Class II
Quantity
2
Official record key
device-enforcement:Z-2097-2017

Official wording

Reason: A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.

Code information: Serial numbers: PAT.003, PAT.006

Distribution pattern: FL

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.