Skip to content
Recall Observatory FDA recall evidence

Device product

AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO¿ is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg).

Z-0715-2018

September 26, 2017

Class II

Product summary

Firm
Mobius Imaging, LLC
Event
Event 79159
Status
Terminated
Classification
Class II
Quantity
105
Official record key
device-enforcement:Z-0715-2018

Official wording

Reason: If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.

Code information: Requested.

Distribution pattern: Worldwide Distribution - US Nationwide in the states of: AZ, CA, CO, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, MT, NC, NY, OH, OR, PA, RI, TN, TX, UT, VA, WA, WV and the countries of Argentina, Australia, Austria, Belgium, Canada, Egypt, France, Germany, India, Italy, Japan, Kazakhstan, Lithuania, Netherlands, Pakistan, Russia, Saudi Arabia, Spain, Switzerland, UAE, Uzebekistan

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.