Peripherally Inserted Central Catheters (PICC), UPN H965750141, Catalog # 75-014 The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring

Z-0714-2017

June 08, 2016

Class II

Product summary

Firm
Angiodynamics Inc. (Navilyst Medical Inc.)
Event
Event 74948
Status
Terminated
Classification
Class II
Quantity
48 cartons (boxes of 5)
Official record key
device-enforcement:Z-0714-2017

Official wording

Reason: There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.

Code information: Lot #: 4941411

Distribution pattern: Nationwide Distribution to CA, FL, GA, HI, LA, MI, MN, MO, NV, OH & TX

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.