MRIdian ViewRay Radiation Therapy System, ViewRay Treatment Planning and Delivery System (also known as the MRIdian¿ System) is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body when radiation treatment is indicated

Z-1309-2016

March 07, 2016

Class II

Product summary

Firm
Viewray Incorporated
Event
Event 73550
Status
Terminated
Classification
Class II
Quantity
5 units
Official record key
device-enforcement:Z-1309-2016

Official wording

Reason: When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a result there is the potential to deliver dose to the initial isocenter rather than the new location.

Code information: Model number: 10000 Catalog/Part Number: 10000 Serial Numbers: 100, 101, 102, 104 & 105

Distribution pattern: Worldwide Distribution - US including CA, FL, MO & WI., and International distribution to the Republic of Korea and The Netherlands.

Derived failure modes