Skip to content
Recall Observatory FDA recall evidence

Device product

Innova Self-Expanding Stent System, 200 mm. Labeled with Material Number and description as follows: Product Usage: The Innova Vascular Self-Expanding Stent System is indicated to improve luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery (PPA) with reference vessel diameters from 4.0 mm to 7.0 mm and lesion lengths up to 190 mm. a. H74939293052030 (Material Description Innova 5 x 200 x 130); b. H74939293052070 (Material Description Innova 5 x 200 x 75); c. H74939293062030 (Material Description Innova 6 x 200 x 130); d. H74939293062070 (Material Description Innova 6 x 200 x 75); e. H74939293072030 (Material Description Innova 7 x 200 x 130); f. H74939293072070 (Material Description Innova 7 x 200 x 75); g. H74939293082030 (Material Description Innova 8 x 200 x 130); h. H74939293082070 (Material Description Innova 8 x 200 x 75)

Z-0685-2018

November 06, 2017

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 78587
Status
Ongoing
Classification
Class II
Quantity
22,916 units total
Official record key
device-enforcement:Z-0685-2018

Official wording

Reason: Stent possibly unable to be fully released from the delivery system.

Code information: a. Innova 5 x 200 x 130 (Batch No. 18095733, 18137885, 18233108, 18379812, 18362996, 18420576, 18404859, 18457350, 18478823, 18488703, 18519986, 18696468, 18670333, 18667082, 18756979, 18855752, 18844729, 18872178, 18886680, 18866631, 18892405, 18900294, 18951278, 19039704, 19064055, 19079251, 19111601, 19173251, 19206070, 19238815, 19231426, 19351863, 19309202, 19382422, 19367647, 19426956, 19483869, 19490837, 19501322, 19540549, 19540559, 19569664, 19641827, 19750808, 19742326, 19830060, 19844547, 19873365, 19907333, 19970577, 20014162, 20080095, 20096498, 20112175, 20186872, 20223208, 20216043, 20435880, 20151952, 20444934, 20439618, 20314154, 20336209, 20394072, 20608020, 20608137, 20607734, 20657170, 20420612, 20337854, 20450027, 20616298, 20777472, 20470018, 20672513, 20832034, 20876897, 20871401, 20728970, 20776633, 20851808, 20975844, 21020806, 21124105, 21045072, 21057925, 21091417, 21115593); b. Innova 5 x 200 x 75 (Batch No. 18343796, 18115802, 18379813, 19013283, 21124110); c. Innova 6 x 200 x 130 (Batch No. 18152540, 18137923, 18096697, 18119044, 18119040, 18115810, 18230863, 18108301, 18230864, 18211280, 18211281, 18245462, 18230637, 18230636, 18410450, 18428643, 18473268, 18283383, 18343799, 18439907, 18431333, 18368130, 18523235, 18519980, 18661526, 18488709, 18389662, 18498563, 18515152, 18446693, 18733063, 18759192, 18759189, 18766678, 19540741, 18791863, 18832544, 18792872, 18855754, 18860236, 18823216, 18866636, 18844738, 18892194, 18900365, 18870948, 18872541, 18878446, 18900381, 18888435, 18988348, 18961203, 19064043, 18945808, 19072771, 19111551, 19079253, 19151922, 19116475, 19164618, 19206072, 19226126, 19141574, 19245284, 19240145, 19288826, 19288832, 19309205, 19309206, 19367648, 19377502, 19408962, 19426948, 19389007, 19490838, 19511466, 19501332, 19517646, 19545788, 19555598, 19560848, 19569663, 19589370, 19584757, 19641826, 19635253, 19664685, 19654162, 19670616, 19709446, 19728413, 19699320, 19782868, 19693686, 19682956, 19682955, 19830064, 19873362, 19907334, 19905446, 19941738, 19942503, 19918783, 20013959, 19970579, 20014164, 20023939, 20109861, 20188996, 20080098, 20080097, 20188052, 20187538, 20080096, 20109863, 20112179, 20320562, 20830984, 20216045, 20183756, 20158213, 20433364, 20434473, 20233669, 20286988, 20233670, 20303729, 20342627, 20233668, 20607288, 20574727, 20607402, 20455108, 20368051, 20432603, 20511193, 20394775, 20594724, 20635843, 20776979, 20478648, 20821884, 20754076, 20758252, 20821886, 20800469, 20886809, 20825353, 20885852, 20938124, 20913156, 20950528, 20975845, 20997424, 21004412, 21119055, 21117395, 21058225, 21098042, 21071017, 21122202, 21145025, 21145024, 21219047, 21219242, 21219052); d. Innova 6 x 200 x 75 (Batch No. 18389661, 18623249, 18870949, 18900366, 19161485, 19245285, 19918786, 20635844, 20862432, 21098043); e. Innova 7 x 200 x 130 (Batch No. 18152541, 18099223, 18130828, 18113523, 18230638, 18217571, 18211282, 18186723, 18240225, 18168041, 18291418, 18339045, 18368242, 18283384, 18339044, 18316737, 18394605, 18385509, 18420569, 18313150, 18488713, 18921110, 18961213, 19072776, 19134318, 19173258, 19161488, 19206078, 19228862, 19255680, 19288830, 19309211, 19377507, 19327253, 19408970, 19436702, 19483875, 19491821, 19501423, 19527344, 19555961, 19545789, 19624864, 19659433, 19693528, 19670619, 19709445, 19830069, 19811429, 19742336, 19918785, 19907336, 19851529, 19974524, 19974523, 20080094, 20001802, 20109860, 20001803, 20188364, 20109862, 20177945, 20181650, 20210393, 20244728, 20321275, 20320380, 20286989, 20434033, 20267930, 20405389, 20381126, 20316570, 20574731, 20607286, 20352590, 20442151, 20466869, 20622461, 20655052, 20495971, 20518171, 20628127, 20777268, 20777261, 20820730, 20731022, 20833824, 20758256, 20886953, 20819295, 20856506, 20910181, 20937554, 21004415, 21119514, 20997428, 21119515, 21045505, 21103020, 21058247, 21138682, 21217652); f. Innova 7 x 200 x 75 (Batch No. 18336507, 18598712, 18733075, 18802238, 18892409, 18999955, 19151414, 19245286, 19511477, 19569668, 19684652, 19811430, 20699739, 20729796, 20978808, 21122996); g. Innova 8 x 200 x 130 (Batch No. 18211294, 18313156, 18385119, 18519981, 18583365, 18611781, 18733076, 18754584, 18780335, 18844724, 18977671, 19006855, 19048092, 19072779, 19188060, 19245289, 19304348, 19351866, 19454525, 19501424, 19624788, 19664694, 19684647, 19839023, 19782933, 20014160, 20080099, 20112342, 20216047, 20286991, 20439603, 20386608, 20655679, 20655677, 20329707, 20410534, 20655798, 20357647, 20997431, 21123358, 21123667); h. Innova 8 x 200 x 75 (Batch No. 18115809, 18313157, 18877978, 19161490, 19624794, 19887625, 20410535, 20862411)

Distribution pattern: Worldwide Distribution - US Nationwide Distributed domestically to AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV. Distributed internationally to Canada

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Stent possibly unable to be fully released from the delivery system.