Device product
Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96 Determinations, IVD
Z-2782-2016
Product summary
- Event
- Event 74932
- Status
- Terminated
- Classification
- Class III
- Quantity
- US: 6 kits, Foreign: 50 kits
- Official record key
device-enforcement:Z-2782-2016
Official wording
Reason: The product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive control to be invalid (running high out of range).
Code information: Lot Number 2326060-043, Expiration Date 31 May 2017
Distribution pattern: Distribution to US nationwide including California, and China (Hong Kong).
Derived failure modes
-
Unknown
The product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive control to be invalid (running high out of range).