AQURE System; Model Number: 933-599. The AQURE System manages blood gas and immunoassay analyzers.

Z-0748-2016

December 29, 2015

Class II

Product summary

Firm
Radiometer America Inc
Event
Event 72959
Status
Terminated
Classification
Class II
Quantity
375
Official record key
device-enforcement:Z-0748-2016

Official wording

Reason: The AQURE System has a design error regarding sample type in which sample type may be specified for some results and left blank for other results. The fact that the sample type field is left blank in some cases could lead to misinterpretation of results.

Code information: Model Number: 933-599 Version 1.8.3

Distribution pattern: Worldwide Distribution-US (nationwide) including the states of PA, GA, MI, OH, VA, FL, AL, TX, OR, OK, TN, NY, and MA, and the countries of Austria, Czech Republic, Denmark, Estonia, United Emirates, Kuwait, Qatar, Saudi Arabia, Oman, Jordan, Finland, France, Germany, Hungry, Ireland, United Kingdom, Italy, Netherlands, Norway, Poland, Spain, Sweden, Switzerland, Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Russia, Singapore, South Africa, Republic of Korea, and Malaysia.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    design error