Device product
BARD Biopsy Systems (A Business Unit of Bard Peripheral Vascular, Inc.), UltraClip¿ Dual Trigger Breast Tissue Marker 17g X 10cm needle, Ultrasound Enhanced Coil Marker 5 Pack; Sterilize, Non-pyrogenic, Rx only, Single Use; Bard Peripheral Vascular, Inc., 1625 West 3rd Street, Tempe, AZ 85281 USA. PK4320100 Rev. 0 03/10 The UltraClip Dual Trigger breast tissue marker is a sterile, single use device comprised of a disposable introducer, and a metal implantable tissue marker with polyvinyl alcohol (PVA).
Z-1526-2013
Product summary
- Event
- Event 64276
- Status
- Terminated
- Classification
- Class II
- Quantity
- 815
- Official record key
device-enforcement:Z-1526-2013
Official wording
Reason: Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard¿ UltraClip¿ Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization process (non sterile).
Code information: PC 864017D; Lots: HUWK1296, HUWK1297
Distribution pattern: Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KY, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, TX, VA, WA, WV and WY.
Derived failure modes
-
Unknown
Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard¿ UltraClip¿ Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization process (non sterile).