Device product
IMPLANT,TM, MTX FULL 3.7 mm x 11.5 mm diameter 3.5 mm Catalog #TMTB11 Product Usage: Zimmer Dental implant systems are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load.
Z-0782-2016
Product summary
- Event
- Event 73013
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1059
- Official record key
device-enforcement:Z-0782-2016
Official wording
Reason: Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
Code information: Lot # 63193920 63244217
Distribution pattern: Worldwide Distribution. US Nationwide in the states of NY, CA, KS, NJ, CO, MI, MD, WA, CT, KY, OR, OH, IA, SC, AZ, AK, IN, TN, WI, FL, AL, UT, WY, MA, PA, MN, IL, MI, GA, ND, RI, including Puerto Rico; and the countries of Thailand, France, Germany, Colombia, Hungary, Netherlands, Saudi Arabia, Poland, Spain, and Italy
Derived failure modes
-
Unknown
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.