Skip to content
Recall Observatory FDA recall evidence

Device product

EnVision FLEX+, Mouse, High pH (Link), Product No. K8002, K8004. Intended for use in immunohistochemistry together with Autostainer Link Instruments.

Z-2330-2016

July 06, 2016

Class II

Product summary

Firm
Dako Denmark A/S
Event
Event 74608
Status
Terminated
Classification
Class II
Quantity
147 kits
Official record key
device-enforcement:Z-2330-2016

Official wording

Reason: The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.

Code information: K8002 Lot no: 10112879 K8004 Lot no: 10111463

Distribution pattern: Worldwide Distribution -- US, Canada, and Colombia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The pH of Target Retrieval Solution in the kit measures 4-5 instead of the correct pH of 9.1.