Skip to content
Recall Observatory FDA recall evidence

Device product

Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye.

Z-1404-2013

December 05, 2012

Class II

Product summary

Firm
Canon Inc.
Event
Event 65082
Status
Terminated
Classification
Class II
Quantity
384 units
Official record key
device-enforcement:Z-1404-2013

Official wording

Reason: Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2s refractive power (SPH value) sometimes indicated abnormal value which occurs approximately 5% in probability.

Code information: Catalog Number 6937B002AA. All serial numbers from the first unit onward.

Distribution pattern: Worldwide Distribution - USA Nationwide and countries of: Australia, Japan, Korea, Singapore and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2s refractive power (SPH value) sometimes indicated abnormal value which occurs approximately 5% in probability.