BD Viper LT software, version 3.00H

Z-1866-2017

December 14, 2015

Class II

Product summary

Firm
Bd Diagnostic
Event
Event 76895
Status
Terminated
Classification
Class II
Quantity
2
Official record key
device-enforcement:Z-1866-2017

Official wording

Reason: BD has confirmed reports that lower than expected RFU values were generated on customer BD Viper LT Systems after the installation of software version 3.00H. As RFU values decrease, results approach the assay cutoff threshold which has the potential to impact test results.

Code information: Catalog Number 443157 443153 443155 443156 442947 443902 Version Number 3.00H 3.00H

Distribution pattern: 2 US customers, 0 Ex-US regions, and 0 distributors 0.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    BD has confirmed reports that lower than expected RFU values were generated on customer BD Viper LT Systems after the installation of software version 3.00H. As RFU values decrease, results approach the assay cutoff threshold which has the potential to impact test results.