Philips Healthcare IQon Spectral CT computed tomography x-ray system
Z-0333-2018
Product summary
- Event
- Event 78879
- Status
- Terminated
- Classification
- Class II
- Quantity
- 50
- Official record key
device-enforcement:Z-0333-2018
Official wording
Reason: There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.
Code information: IQon Spectral CT 728332 Devices manufactured 03/05/2009 to present.
Distribution pattern: 107 domestic accounts 15 US Govt accounts 682 Foreign Accounts 3 Mexico accounts 12 Canada accounts
Derived failure modes
-
Unknown
There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.