RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This device can also be used with the Total Performance System (TPSTM).

Z-1742-2013

July 01, 2013

Class II

Product summary

Firm
Stryker Instruments Div. of Stryker Corporation
Event
Event 65456
Status
Terminated
Classification
Class II
Quantity
9
Official record key
device-enforcement:Z-1742-2013

Official wording

Reason: The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100

Code information: 0910505713, 1029801073, 1225400823, 1314201603, 1314300713, 1314300723, 1314300733, 1314300743, 1314300753, 1314300763, 1314300773, 1314300783, 1314300793, 1314300803, 1314300813, 1314300823, 1314300833, 1314300843, 1314300853, 1314300863, 1314300873, 1314300883, 1314300893, 1314300903, 1314804033, 1314900053, 1314906453, 1314908193

Distribution pattern: Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software. Handpieces that have been programmed with the incorrect