Device product

12-Instrument Sterilization Tray; Product Usage: The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.

Z-1544-2014

April 16, 2014

Class II

Product summary

Firm
Stryker Endoscopy
Event
Event 68045
Status
Terminated
Classification
Class II
Quantity
2591 trays
Official record key
device-enforcement:Z-1544-2014

Official wording

Reason: Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.

Code information: Model number: 0242000012; Lot/serial numbers: 021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1, 48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1, 36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129, 43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476, 43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542, 41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1, 51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1, 37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218., 43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1, 48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445, 42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843, 42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610, 43893-1-13, 39219, 43893-1-2, 43893-1-14.

Distribution pattern: Worldwide distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.