Skip to content
Recall Observatory FDA recall evidence

Device product

iLab Ultrasound Imaging System; iLab 3.0 (Polaris 1.0) software. Used for ultrasound examinations of intravascular pathology.

Z-1487-2014

April 03, 2014

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 67959
Status
Terminated
Classification
Class II
Quantity
9 units
Official record key
device-enforcement:Z-1487-2014

Official wording

Reason: Specific iLab Ultrasound Imaging Systems using newest 3.0 (Polaris 1.0) software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may display an inaccurate value that is double the actual distance.

Code information: Model Numbers: H749ILAB120CART0 H749ILAB120INS0 H749ILAB120N2710 Serial numbers: 4219, 4900, 5701, 5827 7516, 7517, 7607 8407, 8434. the different UPNs represent minor level differences in configuration relative to mobility and no difference in any essential components or software.

Distribution pattern: USA Nationwide Distribution in the states of NY, NC, DC, FL, PA

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software change the pullback speed from the default setting of 0.5 mm/sec to 1.0 mm/sec, the Longview distance measurement may