Device product
A-series lift - A625 Manual w/Load Cell & Wi-Fi, Omni Part No. 324124. Prism Medical.
Z-1405-2017
Product summary
- Event
- Event 76351
- Status
- Terminated
- Classification
- Class II
- Quantity
- 21
- Official record key
device-enforcement:Z-1405-2017
Official wording
Reason: During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.
Code information: A6MC000110 A6M0001010 A6M0001011 A6M0001012 A6M0001013 A6M0001014 A6M0001015 A6M0001016 A6M0001017 A6M0001018 A6M0001019 A6M0001020 A6M0001021 A6M0001025 A6M0001026 A6M0001029 A6M0001022 A6M0001023 A6M0001024 A6M0001027 A6M0001028
Distribution pattern: Product was distributed to MT and NV; one VA consignee.
Derived failure modes
-
Unknown
During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.