Device product
iConnect Access used with Merge PACS software provides medical specialists with access to diagnostic quality images, reports, and various types of patient data over conventional TCP/IP (e.g., internet) networks.
Z-2531-2016
Product summary
- Event
- Event 74734
- Status
- Terminated
- Classification
- Class II
- Quantity
- 146 sites potentially have the affected versions for both products
- Official record key
device-enforcement:Z-2531-2016
Official wording
Reason: Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were ingested in PACS.
Code information: iConnect Access V4.0 or earlier with any version of Merge PACS and iConnect Access V4.1 with Merge PACS V6.5.4 or earlier
Distribution pattern: Distribution was made to medical facilities nationwide and to one foreign medical facility in New Zealand. There was no military or government distribution.
Derived failure modes
-
Unknown
Studies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were ingested in PACS.