REF 5100-088, TPS Micro Driver, For Use with Stryker Total Performance System Caution Federal Law (USA) restricts the device to use by or on the order of a physician Driver, Wire and Bone Drill; The Stryker Total Performance System (TPS) is intended for use in the cutting, drilling, reaming, decorticating and smoothing of bone and other bone related tissues for a variety of applications such as ENT, dental, orthopedic, maxillofacial, spinal, and plastic surgery. The instruments are also used in the placement of screws, wires, pins, and other fixation devices.

Z-0092-2014

October 02, 2013

Class II

Product summary

Firm
Stryker Instruments Div. of Stryker Corporation
Event
Event 66352
Status
Terminated
Classification
Class II
Quantity
67
Official record key
device-enforcement:Z-0092-2014

Official wording

Reason: It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma

Code information: 10933, 11433, 11483, 20553, 30163, 31553, 40713, 40833, 41113, 50463, 60133, 60423, 90383, 110093, 110193, 1030133, 1030403, 1040413, 2010013, 2010123, 96090403, 96091343, 97020193, 97030013, 97030543, 97050243, 97060163, 97080033, 98010083, 98020703, 98030103, 98040213, 98060023, 98061033, 98070603, 98090023, 98090473, 98090503, 98101143, 98110323, 99010433, 99030253, 99050033, 99050383, 99060163, 99060423, 99061063, 99061283, 99070453, 99071293, 99080243, 99080483, 99090913, 99101273, 221303733, 326102033, 434909453, 500800103, 504900233, 631112793, 631414853, 700314923, 723606673, 725315933, 735502183, 813304543, and 921700523 ,

Distribution pattern: Worldwide distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Italy, Japan, Spain, Switzerland, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It was discovered that the silicone potting failed to properly cure after undergoing the prescribed curing possess. The potting material is used to insulate, stabilize and provide a moisture barrier to the TPS MicroDriver main Printed Circuit Board Assembly (PCBA). Inadequate cure of the silicone potting could potentially allow it to leak to the outside of the handpiece or cause an electrical ma