Infant Warmer System (IWS)

Z-2323-2017

May 05, 2017

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 77287
Status
Terminated
Classification
Class II
Quantity
34986
Official record key
device-enforcement:Z-2323-2017

Official wording

Reason: GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

Code information: Medical device listing number: D103053, D103054, D103056, D103057, D183529

Distribution pattern: Worldwide Distribution - USA (nationwide) Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.