Device product
ADVIA Chemistry Urea Nitrogen Reagent (material number 10309051) for use on ADVIA 1800, 2400 and XPT Chemistry Systems
Z-2062-2018
Product summary
- Event
- Event 79932
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-2062-2018
Official wording
Reason: Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails.
Code information: Lot # Expiry UDI# 408143 2018-06-28 00630414517513 418290 2018-06 - 28 00630414517513
Distribution pattern: Nationwide and worldwide
Derived failure modes
-
Unknown