Device product
ECT Internal Fracture Fixation Quick-Connect Drill 2.5 mm Diameter 11 cm Length, Item Number/EDI 00241002500, Nonsterile. orthopedic surgical instrument.
Z-3184-2017
Product summary
- Event
- Event 77417
- Status
- Terminated
- Classification
- Class II
- Quantity
- 412 units
- Official record key
device-enforcement:Z-3184-2017
Official wording
Reason: Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
Code information: Lot Numbers: 63289918, 63296044, 63451593, 63457958, 63462351, 63469945, 63479047 & 63518080.
Distribution pattern: Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
Derived failure modes
-
Unknown
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.