Skip to content
Recall Observatory FDA recall evidence

Device product

Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Z-1930-2018

February 21, 2018

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc.
Event
Event 80071
Status
Terminated
Classification
Class II
Quantity
14 units
Official record key
device-enforcement:Z-1930-2018

Official wording

Reason: Possibility that the Biograph Horizon systems performing CT retrospective cardiac gating or PET cardiac gating examinations may experience a waveform sampling issue caused by a firmware change within the Universal Physiological Measurement Module.

Code information: Code No. 10532746, 10532748

Distribution pattern: Domestic distribution to . International distribution to Finland, France, France, Iran, Ireland, Italy, Japan, Japan, Netherlands.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    issue caused by a firmware