Device product
NucliSENS Lysis Buffer Extended lot for RES 76675
Z-0086-2018
Product summary
- Firm
- BioMerieux SA
- Event
- Event 78387
- Status
- Terminated
- Classification
- Class II
- Quantity
- 911 units
- Official record key
device-enforcement:Z-0086-2018
Official wording
Reason: Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR reactions and result in an uninterpretable test result. If using an Internal Control (IC) in the extraction process the IC could potentially be inhibited as well and invalidate test results.
Code information: REF 200292, Lot 17022802, Exp Date 01/28/2019 Extended lot for RES 76675
Distribution pattern: Distributed domestically to MD, CO, and OH.
Derived failure modes
-
Unknown
Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR reactions and result in an uninterpretable test result. If using an Internal Control (IC) in the extraction process the IC could potentially be inhibited as well and invalidate test results.