Device product
K-Assay(R) Multi-Analyte Calibrator, REF KAI--016C, KAMIYA BIOMEDICAL COMPANY, Seattle, WA, IVD The K-ASSAY(R) Multi-Analyte Calibrator is intended to be used for the calibration of the K-ASSAY(R) Alpha-1 Acid Glycoprotein, Alpha-1 Anti-Trypsin, Complement C3, Complement C4, Haptoglobin, lgA, lgG, lgM, and Transferrin immunoturbidimetric assays. For in vitro diagnostic use
Z-3138-2017
Product summary
- Event
- Event 77916
- Status
- Terminated
- Classification
- Class II
- Quantity
- 1398 units
- Official record key
device-enforcement:Z-3138-2017
Official wording
Reason: The AlAT calibrator values in the package insert for the Multi-Analyte Calibrator, lot D755, lots N653 and lot N654 were not correct.
Code information: Lot Numbers: N653, N654, D755, D647, F648, F649, G650, J651, L652
Distribution pattern: Worldwide Distribution - US including FL, CA, WA, OH, IA, IL, NJ, NY, NC, MA, TX, CT, MD, VA, Internationally to TN, Czech Republic, Taiwan, Dominican Republic, Switzerland, China, and Jamaica
Derived failure modes
-
Unknown
The AlAT calibrator values in the package insert for the Multi-Analyte Calibrator, lot D755, lots N653 and lot N654 were not correct.