Skip to content
Recall Observatory FDA recall evidence

Device product

Tenor is a mobile passive lift, intended to be used on horizontal surfaces for lifting and transfer in hospitals, nursing homes or other healthcare facilities. Transfers with the Tenor are made indoors in the patient's room, in communal areas or in a bathroom. Tenor is meant to be used with the patient type who: sits in a wheelchair, has no capability to support themselves at all, cannot stand unsupported and is not able to bear weight (not even partially), is dependent on caregiver in most situations, is a passive resident, might be almost completely bed ridden, is often stiff and has contracted joints, and/or is totally dependent. The Tenor has a safe working load of 320 KG (705 lbs).

Z-2652-2014

June 10, 2014

Class II

Product summary

Firm
ARJOHUNTLEIGH POLSKA SP. ZO.O.
Event
Event 68113
Status
Terminated
Classification
Class II
Quantity
382 lifts total worldwide (160 US)
Official record key
device-enforcement:Z-2652-2014

Official wording

Reason: ArjoHuntleigh received reports where the Tenor Lift without Scale (spreader bar) bolt broke during use. There is a remote risk concerning the breakage of the spreader bar bolt connection, part # TEN.018. During use, the bolt is subjected to mechanical forces, which may lead to breakage.

Code information: Model Number: KHA1000-US; Serial Numbers: P0111930 , P0112213 , P0112214 , P0117472 , P0121496 , P0123586 , P0125147 , P0127106 , P0127347 , P0127348 , P0128923 , P0128924 , P0129153 , P0129154 , P0129369 , P0129370 , P0131041 , P0131526 , P0131527 , P0131773 , P0131774 , P0133024 , P0134145 , P0134146 , P0136335 , P0137220 , P0138564 , P0139494 , P0139495 , P0139696 , P0144387 , P0148499 , P0149211 , P0149302 , P0149303 , P0162196 , P0164314 , P0165868 , P0165941 , P0166383 , P0166464 , P0167748 , P0170439 , P0172605 , P0172717 , P0179187 , P0179414 , P0180046 , P0180047 , P0180480 , P0180481 , P0180482 , P0180483 , P0183127 , P0183128 , P0183764 , P0183765 , P0183824 , P0184254 , P0186224 , P0186225 , P0187652 , P0187653 , P0188944 , P0189050 , P0189099 , P0189762 , P0192050 , P0192249 , P0208300 , P0209851 , P0210756 , P0210792 , P0210793 , P0211855 , P0211856 , P0211980 , P0211981 , P0212717 , P0217417 , P0218714 , P0220790 , P0221243 , P0221244 , TCMA0405021 , TCMNS0404E154 , TCMNS0404E155 , TCMNS0404U044 , TCMNS0404U045 , TCMNS0408E159 , TCMNS0408E160 , TCMNS0408E161 , TCMNS0409E163 , TCMNS0409G144 , TCMNS0409G145 , TCMNS0409G146 , TCMNS0409G147 , TCMNS0409U046 , TCMNS0409U047 , TCMNS0409U048 , TCMNS0409U049 , TCMNS0409U050 , TCMNS0409U051 , TCMNS0409U052 , TCMNS0409U053 , TCMNS0410G164 , TCMNS0410G165 , TCMNS0410G166 , TCMNS0410G167 , TCMNS0410G168 , TCMNS0410G169 , TCMNS0411E170 , TCMNS0412E228 , TCMNS0412E229 , TCMNS0412G171 , TCMNS0412G172 , TCMNS0412G173 , TCMNS0501G233 , TCMNS0501G234 , TCMNS0501G235 , TCMNS0501U264 , TCMNS0501U265 , TCMNS0501U274 , TCMNS0501U275 , TCMNS0501U276 , TCMNS0502A279 , TCMNS0502E288 , TCMNS0502G266 , TCMNS0502G267 , TCMNS0502G268 , TCMNS0502G269 , TCMNS0502G270 , TCMNS0502G271 , TCMNS0502U320 , TCMNS0502U321 , TCMNS0504U328 , TCMNS0504U329 , TCMNS0504U330 , TCMNS0505E333 , TCMNS0506E350 , TCMNS0507A396 , TCMNS0507A397 , TCMNS0507A398 , TCMNS0508E400 , TCMNS0508E401 , TCMNS0508E402 , TCMNS0508E403 , TCMNS0509U404 , TCMNS0509U405 , TCMNS0510E408 , TCMNS0510U407 , TCMNS0511E410 , TCMNS0511E419 , TCMNS0511E422 , TCMNS0511U409 , TCMNS0511U417 , TCMNS0511U418 , TCMNS0601E454 , TCMNS0602E470 , TCMNS0602E471 , TCMNS0602G472 , TCMNS0602G473 , TCMNS0603U564 , TCMNS0603U565 , TCMNS0603U566 , TCMNS0604G640 , TCMNS0605E641 , TCMNS0606G643 , TCMNS0606G644 , TCMNS0607A664 , TCMNS0607A665 , TCMNS0607A666 , TCMNS0607E658 , TCMNS0607G650 , TCMNS0607G651 , TCMNS0607G656 , TCMNS0607G657 , TCMNS0608E667 , TCMNS0608U669 , TCMNS0608U670 , TCMNS0608U671 , TCMNS0608U699 , TCMNS0608U700 , TCMNS0608U701 , TCMNS0609U702 , TCMNS0609U703 , TCMNS0609U704 , TCMNS0610E736 , TCMNS0610E737 , TCMNS0610E738 , TCMNS0610U708 , TCMNS0610U709 , TCMNS0610U734 , TCMNS0610U735 , TCMNS0612E755 , TCMNS0612U756 , TCMNS0612U757 , TCMNS0701E771 , TCMNS0701E772 , TCMNS0701U774 , TCMNS0703E792 , TCMNS0703E793 , TCMNS0704U824 , TCMNS0704U825 , TCMNS0704U826 , TCMNS0705E828 , TCMNS0705E829 , TCMNS0705G827 , TCMNS0706A868 , TCMNS0706A869 , TCMNS0706G865 , TCMNS0706G866 , TCMNS0706U832 , TCMNS0706U847 , TCMNS0707E888 , TCMNS0707E890 , TCMNS0707E892 , TCMNS0707G889 , TCMNS0707U872 , TCMNS0707U873 , TCMNS0707U891 , TCMNS0708U894 , TCMNS0709U909 , TCMNS0709U910 , TCMNS0709U911 , TCMNS0710E925 , TCMNS0710U940 , TCMNS0710U955 , TCMNS0711E958 , TCMNS0802E020 , TCMNS0802E028 , TCMNS0802E029 , TCMNS0803E056 , TCMNS0803U057 , TCMNS0803U058 , TCMNS0804E072 , TCMNS0804U075 , TCMNS0805A076 , TCMNS0805A077 , TCMNS0805A078 , TCMNS0805A079 , TCMNS0805A080 , TCMNS0805A081 , TCMNS0805A082 , TCMNS0805A083 , TCMNS0805A084 , TCMNS0805A085 , TCMNS0805U093 , TCMNS0805U094 , TCMNS0806G120 , TCMNS0806G121 , TCMNS0806G122 , TCMNS0806G123 , TCMNS0806G124 , TCMNS0807A143 , TCMNS0808E160 , TCMNS0808E163 , TCMNS0808U161 , TCMNS0808U162 , TCMNS0808U165 , TCMNS0808U166 , TCMNS0808U167 , TCMNS0808U168 , TCMNS0809E169 , TCMNS0809E171 , TCMNS0809U170 , TCMNS0810E188 , TCMNS0810E205 , TCMNS0810E206 , TCMNS0810E207 , TCMNS0810E209 , TCMNS0810U190 , TCMNS0810U191 , TCMNS0810U192 , TCMNS0810U193 , TCMNS0811E225 , TCMNS0811E226 , TCMNS0811E227 , TCMNS0811E228 , TCMNS0901E232 , TCMNS0901E271 , TCMNS0901U266 , TCMNS0901U267 , TCMNS0901U268 , TCMNS0901U269 , TCMNS0902G313 , TCMNS0902G314 , TCMNS0902U296 , TCMNS0902U302 , TCMNS0903G318 , TCMNS0903G319 , TCMNS0904E321 , TCMNS0904G322 , TCMNS0904U338 , TCMNS0904U339 , TCMNS0904U367 , TCMNS0904U368 , TCMNS0904U369 , TCMNS0905G372 , TCMNS0905G373 , TCMNS0905G374 , TCMNS0905G375 , TCMNS0905G376 , TCMNS0905G377 , TCMNS0905U370 , TCMNS0905U389 , TCMNS0905U390 , TCMNS0905U391 , TCMNS0905U392 , TCMNS0907U411 , TCMNS0907U412 , TCMNS0907U413 , TCMNS0907U414 , TCMNS0908E441 , TCMNS0908U426 , TCMNS0908U427 , TCMNS0908U428 , TCMNS0908U429 , TCMNS0910U445 , TCMNS0910U446 , TCMNS0910U447 , TCMNS0910U448 , TCMNS0910U449 , TCMNS0910U450 , TCMNS0910U451 , TCMNS0910U452 , TN0911E003 , TN0911E004 , TN0911E005 , TN1002G006 , TN1004U007 , TN1004U008 , TN1005E011 , TN1005U009 , TN1005U010 , TN1006U012 , TN1006U013 , TN1009E016 , TN1009U014 , TN1009U015 , TN1009U017 , TN1010U018 , TN1010U019 , TN1010U020 , TN1010U021 , TN1012E023 , TN1012U022 , TN1102E001 , TN1102E002 , TN1103G003 , TN1103G004 , TN1103G005 , TN1103G006 , TN1103G007 , TN1103G008 , TN1103G009 , TN1104E016 , TN1104E017 , TN1104E018 , TN1104U010 , TN1104U011 , TN1104U012 , TN1104U013 , TN1104U014 , TN1104U015 , TN1105E021 , TN1105U019 , TN1105U020 , TN1106E022 , TN1106E023 , TN1106E024 , TN1106U025 , TN1106U026 , TN1106U027 , TN1106U028 , TN1106U029 , TN1107E030 , TN1107E031 , TN1107E032 , TS0911E001 , TS0911E002 , TS1108A033

Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to Australia, Bahrain, Canada, France, Germany, India, Ireland, Israel, Italy, Mexico, Netherlands, New Zealand, Poland, Romania, Saudi Arabia, South Africa, Spain, Sweden, United Kingdom, and Uruguay.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    ArjoHuntleigh received reports where the Tenor Lift without Scale (spreader bar) bolt broke during use. There is a remote risk concerning the breakage of the spreader bar bolt connection, part # TEN.018. During use, the bolt is subjected to mechanical forces, which may lead to breakage.