Device product
ADVIA Centaur Calibrator 30, 2 pack, 6 pack, and Ref 2 pack. Siemens Material Number (SMN): 10379810, 10379811, 10697756.
Z-0515-2018
Product summary
- Event
- Event 78645
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5992 units (US and OUS)
- Official record key
device-enforcement:Z-0515-2018
Official wording
Reason: Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).
Code information: Lot Numbers: 00387A21, 16673A21, 36243A21; 00388A21, 15448A21, 51211A21; 20640A21; 51024A21. Expiration Date: 2018-03-08
Distribution pattern: US and OUS.
Derived failure modes
-
Unknown
Siemens Healthcare Diagnostics has confirmed a negative bias for ADVIA Centaur Enhanced Estradiol (eE2) on the ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur XPT systems when calibrating with Calibrator 30 kit lots ending in 21 (C3021) as compared to Calibrator 30 kit lots ending in 20 (C3020).