Device product
LifeScan brand OneTouch¿ Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing of Blood Glucose Levels.
Z-1129-2013
Product summary
- Firm
- Lifescan Inc
- Event
- Event 64478
- Status
- Terminated
- Classification
- Class II
- Quantity
- 132,301 units total
- Official record key
device-enforcement:Z-1129-2013
Official wording
Reason: When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.
Code information: 2AA2G01, Exp. Date: 7-31-2014; 2AA2G02, Exp. Date: 7-31-2014; 2AA2G03, Exp. Date: 8-31-2014; 3AA2G04, Exp. Date: 12-31-2014; 3AA2G05, Exp. Date: 12-31-2014
Distribution pattern: Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, 10, IL, IN, KS, KY, LA, MA, MD, ME, Ml, MN, MO, MS, MT, NC, NO, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SO, TN, TX, UT, VA, VI, VT, WA, WI,WV, and WY, and the countries of Russia and Ukraine.
Derived failure modes
-
Unknown