Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 41MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 44MM DIA, ENDO FEMORAL HEAD 45MM DIA, ENDO FEMORAL HEAD 46MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 49MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 51MM DIA, ENDO FEMORAL HEAD 52MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 55MM DIA

Z-2548-2014

July 22, 2014

Class II

Product summary

Firm
Zimmer, Inc.
Event
Event 68818
Status
Terminated
Classification
Class II
Quantity
972,125 units
Official record key
device-enforcement:Z-2548-2014

Official wording

Reason: Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Code information: Item Numbers : 00781804300, 00781805300, 00781804700, 00781805000, 00781804200, 00781805400, 00781804800, 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500 Lot Numbers: 37210958, 37210967, 37211180, 37211183, 37211311, 37211316, 37211320, 37211379, 37211482, 62561484, 37211643, 37211645, 37211646, 37211647, 37211653, 37211655, 37211657, 37211658, 37211661, 37211801, 37211804, 37211805, 37211806, 37211811, 37211812, 37211815, 37211817, 37211820, 37212062, 37212063, 37212065, 37212074, 37212079, 62573574, 62573576, 62576002, 62576008, 62593298, 62593302, 37212608, 37211644

Distribution pattern: Worldwide Distribution.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    design verification test failures