Device product

Intellivue iX Information Center Software - All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0

Z-0776-2018

January 03, 2018

Class II

Product summary

Firm
Philips Electronics North America Corporation
Event
Event 79132
Status
Terminated
Classification
Class II
Quantity
20357
Official record key
device-enforcement:Z-0776-2018

Official wording

Reason: Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.

Code information: All PIIC iX Surveillance stations including: 866023 IntelliVue Info Center iX A.0 866117 PIIC Classic Upgrade 866389 IntelliVue Info Center iX B.0, C.0 867141 IntelliVue Info Center iX B.0 866025 IntelliVue Server IX A.0 866118 Database Server Upgrade A.0

Distribution pattern: Domestic and Foreign

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.