da Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon in the accurate control of surgical instruments such as endoscopes, scissors, forceps, and retractors during endoscopic surgery. The system has four universal manipulators (arms) which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures in the pelvic, abdominal, and thoracic cavities.

Z-2630-2017

June 01, 2017

Class II

Product summary

Firm
Intuitive Surgical, Inc.
Event
Event 77389
Status
Terminated
Classification
Class II
Quantity
10 units
Official record key
device-enforcement:Z-2630-2017

Official wording

Reason: da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially under-torqued brake screws in a section of the system arms.

Code information: Serial Numbers: SK0822, SK1154, SK1001, SK1170, SK1040, SK1176, SK1082, SK1180, SK1146, SK1190

Distribution pattern: Worldwide Distribution - US Nationwide in the states of TX, NY,FL, IL, OH, and countries of France, Ireland, Japan, and Taiwan

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially under-torqued brake screws in a section of the system arms.