Device product
Auriga XL 4007 Laser System, Universal Part Number: M0068FS4007G0. Intended to be used in surgical procedures.
Z-0898-2018
Product summary
- Event
- Event 79096
- Status
- Terminated
- Classification
- Class II
- Quantity
- 9
- Official record key
device-enforcement:Z-0898-2018
Official wording
Reason: Certain Auriga consoles have incorrectly assembled trigger wire components. The trigger wire functions in the start-up of the console. An incorrectly assembled trigger wire may result in error codes or failure to start the console.
Code information: Serial Number: 20509715, 20465096, 20503137, 20520960, 20541967, 20546219, 20546440, 20575512, and 20575518.
Distribution pattern: International Distribution Only -- Australia, Vietnam, France, Germany, Italy, Spain, and Romania.
Derived failure modes
-
Unknown
Certain Auriga consoles have incorrectly assembled trigger wire components. The trigger wire functions in the start-up of the console. An incorrectly assembled trigger wire may result in error codes or failure to start the console.